BD recalls selected batches of BD Connecta three-way stopcocks
The manufacturer, BD Switzerland Sàrl, is withdrawing from distribution and use selected batches of stopcocks from the BD Connecta line. They were placed on the market between spring 2025 and September 2026. The reason is the risk of the tubing disconnecting from the male luer connector.

© Urząd Rejestracji Produktów Leczniczych, Wyrobów Medycznych i Produktów Biobójczych
- 0views
- 0likes and saves
- 0skips
In brief
- BD Switzerland is withdrawing from sale and use selected lot numbers of BD Connecta three-way stopcocks.
- The affected products were distributed from spring 2025 to September 2026.
- The reason is the risk of the tubing detaching from the male luer connector.
What the withdrawal covers
Safety notice number MDS-26-06177 covers specific lot numbers of three-way stopcocks from the BD Connecta line. These products were distributed between spring 2025 and September 2026. The manufacturer's decision applies both to sales and to use in healthcare facilities.
Reason for the decision
As the notice explains, the problem is the risk of the tubing detaching from the male luer connector. During the administration of fluids or medications, this could interrupt therapy or lead to an uncontrolled leak, which is why the manufacturer decided to immediately restrict the use of the affected lots.
Consequences for healthcare facilities
The withdrawal covers the trading and use of the specified lots. Products with the listed lot numbers should no longer be used in healthcare facilities. The manufacturer provides detailed guidance for specific units in the notice itself.
Source of the announcement
The announcement was published by the Office for Registration of Medicinal Products, Medical Devices and Biocidal Products on 17 November 2026.